Gas Chromatography (GC9790II)
Capillary and packed column inlets, split and splitless, cool-on-column
Programmable oven, 8 ramps temperature programming, 6 temperatures, up to 400º C
Wide dynamic range of popular detectors: FID, TCD, NPD, ECD and FPD
Instrument control, method setup and run from keyboard entries
Rugged manual injection and manual pneumatics
Self-diagnosis of controls: Thermal zone, flows, and detectors
Ready to use when product delivered and installed
Unbeatable price and service package
Guarantee timely and no-hassle service responding
Dual pneumatics
| Dual packed column inlets |
| GC9790IIF |
Dual FID Detector |
| GC9790IIT |
TCD Detector |
| GC979IIT-2 |
TCD Detector (new type) |
| GC9790IIP |
FPD Detector |
| GC9790IIE |
ECD Detector |
| GC9790IIG |
Dual FID Detector, dual amplified board |
| GC9790IIB |
Dual FID Detector, TCD Detector |
| GC9790IIB-2 |
Dual FID Detector, TCD Detector (new type) |
| GC9790IID |
FID Detector, FPD Detector |
| GC9790IIK |
Dual FID Detector, ECD Detector |
| GC9790IIH |
TCD Detector, FPD Detector |
| GC9790IIL-3 |
Dual FID Detector, ECD Detector, FPD Detector |
| Packed column inlet, Capillary column inlet (split and splitless) |
| GC9790IIJ |
Dual FID Detector |
| GC9790IIR |
FID Detector, NPD Detector |
| GC9790IIC |
Dual FID Detector, dual FID amplified board |
| GC9790IIA |
Dual FID Detector, TCD Detector |
| GC9790IIA-2 |
Dual FID Detector, TCD Detector (new type) |
| GC9790III |
FID Detector, FPD Detector |
| GC9790IIL |
Dual FID Detector, ECD Detector |
| GC9790IIL-2 |
FID Detector, ECD Detector, FPD Detector |
| GC9790IIL-4 |
FID Detector, ECD Detector, NPD Detector |
| GC9790IIL-5 |
FID Detector, NPD Detector |
| GC9790IIL-6 |
FID Detector, ECD Detector, NPD Detector |
| Dual Capillary column inlets (split and splitless) |
| GC9790IIQ |
Dual FID Detector |
| GC9790IIN |
Dual FID Detector, dual amplified board |
Detector: FID
| Detection limit |
≤3x10 -12 g/s |
| Sensitivity Mv.ml/mg[n-C16] |
|
| Noise |
≤3x10 -14 A |
| Drift |
≤5x10 -13 A/30min |
| Linear range |
≥10 6 |
Detector: TCD
| Detection limit |
|
| Sensitivity Mv.ml/mg[n-C16] |
≥2500 |
| Noise |
≤20μV |
| Drift |
≤100μV/30min |
| Linear range |
≥10 4 |
Detector: TCD(new type)
| Detection limit |
|
| Sensitivity Mv.ml/mg[n-C16] |
≥8000 |
| Noise |
≤20μV |
| Drift |
≤100μV/30min |
| Linear range |
≥10 4 |
Detector: ECD
| Detection limit |
≤1x10 -13 g/ml |
| Sensitivity Mv.ml/mg[n-C16] |
|
| Noise |
≤20μV/30min |
| Drift |
≤50μV/30min |
| Linear range |
≥10 4 |
Detector: FPD
| Detection limit |
Mt≤1.4x10 -12 g/s(P), Mt≤5x10 -11 g/s(S) methyl parathion |
| Sensitivity Mv.ml/mg[n-C16] |
|
| Noise |
≤4x10 -10 A |
| Drift |
≤4x10 -10 A/30min |
| Linear range |
≥10 3 (P), ≥10 2 (S) |
Detector: NPD
| Detection limit |
|
| Sensitivity Mv.ml/mg[n-C16] |
Mt≤5x10 -13 g/s(N) Malathion, Mt≤1x10 -12 g/s(P) azobenzol |
| Noise |
≤4x10 -13 A/30min |
| Drift |
≤2x10 -12 A/30min |
| Linear range |
≥10 3 |