Main Specification
|
Wavelength range
|
190-900nm
|
|
Wavelength accuracy
|
±0.25nm
|
|
Resolution
|
Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30%.
|
|
Baseline stability
|
≤0.004A/30min
|
|
Background correction
|
The D2 lamp background correction capability at 1A is better than 30 times. The S-H background correction capability at 1.8A is better than 30 times.
|
|
Hollow Cathode Lamps
|
Lamp turret
|
6-lamp turret, fully automated scan and peak-picking.
|
|
Lamp current adjustment
|
Automatic adjustment and display. Wide pulse current: 0~25mA, Narrow pulse current: 0~10mA
|
|
Lamp power supply mode
|
Powered by two pulses simultaneously
|
|
Optical System
|
Monochomator
|
Single beam, Czerny-Turner design grating monochromator
|
|
Grating
|
1800 l/mm
|
|
Focal length
|
277mm
|
|
Blazed Wavelength
|
250nm
|
|
Spectral Bandwidth
|
0.1nm, 0.2nm, 0.4nm, 1.2nm automatic change.
|
|
Atomizer
|
Flame/Graphite
|
Flame & Graphite Furnace atomizer automatic changeover
|
|
Detection and Data Processing System
|
Detector
|
Photomultiplier with high sensitivity and wide spectral range.
|
|
Software
|
Ms-Windows 98/Me operating system
|
|
Analytical method
|
Working Curve Method (1st order and 2nd order fitting), Sensitivity auto-correction, Standard Addition Method.
|
|
Concentration
|
Auto-calculation of concentration and content
|
|
Repeatable measurement
|
Maximum 20 times of repeat measurement, automatic calculation of mean value, standard deviation and relative standard deviation.
|
|
Multi-task function
|
Sequential measurement of multi-elements in one sample
|
|
Condition reading
|
With model function
|
|
Result printing
|
Measurement data and final analytical report printout, editing with Excel.
|
|
Flame Analysis
|
Burner
|
10cm single slot all-titanium burner
|
|
Spray chamber
|
All-plastic spray chamber.
|
|
Nebulizer
|
High efficiency glass nebulizer with metal sleeve.
|
|
Position adjustment
|
Adjusting mechanism for vertical, horizontal positions and the rotation angle of the burner.
|
|
Protection for gas system
|
Fuel gas leak protection, abnormal flame extinction alarm, air-pressure deficiency protection and explosion-proof device.
|
|
Flame ignition
|
Automatic
|
|
Characteristic Concentration and Detection Limit
|
Normal Air-C2H2 flame
|
Cu: Characteristic concentration≤0.025mg/L, Detection limit≤0.006mg/L;
|
|
Oxygen-rich Air-C2H2 flame
|
Ba: Characteristic concentration ≤ 0.22mg/L Al: Characteristic concentration ≤ 0.4mg/L
|
|
Hydride generation
|
Hg: detection limit≤0.5ug/L
|
|
Graphite Furnace Analysis
|
Temperature range
|
20~3000ºC
|
|
Light-controlled temperature range
|
1500ºC -3000ºC, temperature auto-correction
|
|
Temperature control accuracy
|
≤1%
|
|
Heating rate
|
2000/s
|
|
Characteristic mass
|
Cd≤0.5 ×10-12g, Cu≤5 ×10-12g, Mo≤1×10-11g
|
|
Precision
|
RSD≤4%
|
|
Safety protection
|
Protections for over-heating, over-current, and less cooling water and argon pressure monitoring functions
|
|
Power Supply
|
220V, 200W
|
|
Dimensions
|
1020 (L) × 492 (W) × 537 (H) mm (main unit)
|
|
420 (L) × 420 (W) × 460 (H) mm (graphite furnace)
|
|
Weight
|
Unpacked 80kg (main unit), 50kg (graphite furnace)
|